Quote:
Originally Posted by Notimetolooz
The problem is that on the lowest cap range the eye stays open without anything connected or a small cap regardless of the capacitance dial setting. I just noticed that if a put a large enough cap on the measuring terminals the eye will close.
|
Did you check the the 200 uuf cap that is used in the lowest range? I see that you have replaced the 2 uf and the .02 uf caps, but apparently not the 200 uuf cap. If it is *open* or perhaps poorly soldered, the C-3 will perform as you described. I just un-soldered the 200uuf cap in my C-3 to verify.